Bently Nevada vb6: Four Channels, 111M7155 and Certification

Industrial machine control panel displaying measurement data during route-based condition monitoring

What is the short answer (TL;DR)?

The Bently Nevada vb6 is a four-channel portable vibration data collector, not a dual-channel instrument. It samples four channels simultaneously through a 24-bit converter at up to 40 kHz Fmax. 111M7155 is the datasheet document number, the instrument order code is VB6-01, and a metrology certificate is a paid option requested when the order is placed.

  • Four channels, not two. The vb6 samples all four inputs at once. Dual-channel operation is a measurement mode capped at 6,400 lines; single-channel spectra reach 12,800 lines and four-channel spectra 3,200 lines.
  • 111M7155 is a document number. It identifies the vb6 datasheet, currently Rev. K. The orderable instrument code is VB6-01, where 01 is the CSA Class I Division 2 agency approval.
  • Software is not in the box. The basic kit carries the instrument, sensors, cables and carry bag. System 1 (3071/01) and Ascent Level 1 and Level 2 (108M4051 and 108M4052) are separate purchases.
  • Free test data is not a calibration certificate. Standard test data can be requested free of charge, but the datasheet says it is not evidence of calibration. Metrology certification is charged and must be requested when the order is placed.
  • Several certification details are unpublished. The datasheet does not name ISO/IEC 17025, the issuing laboratory, a certificate validity period or a metrology order code.

The distinction matters at the quotation stage. A four-channel collector, a software licence and a calibration document are three separate commercial items, and only the first is covered by the basic-kit description.

Does the vb6 collect two channels or four?

The vb6 is a four-channel instrument: all four inputs sample simultaneously, with 24-bit conversion and up to 40 kHz Fmax. Dual-channel operation is a measurement mode, not the hardware limit. Bently Nevada caps dual-channel spectra at 6,400 lines, single-channel spectra at 12,800 lines, and four-channel spectra at 3,200 lines.

The confusion is understandable because the resolution table mentions dual-channel measurements. That row describes what happens when two channels are recorded together, not how many inputs the instrument has. The datasheet description is explicit: the vb6 is “a portable four channel, high resolution vibration data collector” and it “samples all four channels simultaneously.”

  • Four-channel recording is the default capability, useful for triaxial sensors plus a tachometer or a second measurement point.
  • Dual-channel mode keeps more spectral lines per channel, which suits orbit or phase-related work where two channels are enough.
  • Single-channel mode gives the highest resolution, up to 12,800 lines, for detailed spectral diagnosis.
  • Triaxial support is listed as a feature, so three orthogonal axes can be captured in one recording.

For a route-based programme, the practical question is not “two or four channels” but how many measurements must be taken at each bearing without moving the sensor. A four-channel instrument can capture a triaxial reading plus a tachometer input in one pass, while a two-channel device needs more setups.

What is 111M7155, and how is a vb6 ordered?

111M7155 is the document number of the vb6 datasheet, currently Rev. K, not an orderable instrument or spare-part number. The instrument ordering block uses the code VB6-AA, where A is agency approval and option 01 means CSA Class I Division 2. Resolved in distributor catalogues, that code appears as VB6-01.

Bently Nevada uses numeric document numbers for datasheets and manuals, and separate alphanumeric codes for hardware. That is why 111M7155 cannot be entered as a spare part: it will resolve to a PDF, not to an instrument. The actual instrument code is short because the vb6 is sold as a configured portable kit rather than a modular rack assembly.

The ordering block defines one variable field:

  • Agency Approval A: option 01 is CSA, Class I, Division 2, Groups A, B, C and D. No other approval option is listed for the vb6 in the current datasheet.

The rest of the configuration is expressed through the basic-kit contents and the separate software and accessory part numbers, not through a long model string. That is unusual for Bently Nevada, whose rack modules normally carry multi-field order codes, and it is a common source of quotation errors.

vb6 kit and ordering fields at a glance

Field Part number in the datasheet Included in the basic kit?
Instrument VB6-AA template; resolved as VB6-01 (approval 01, CSA) Yes, quantity 1
Straight accelerometers ACCL0547 or 200350 (regional alternates), quantity 3 Yes
Right-angle or straight accelerometer ACCL0561 or 200350 (regional alternates), quantity 1 Yes
Transducer cable 138M7748, 4 ft straight, quantity 4 Yes
Magnetic base MAGF0104, quantity 4 Yes
Signal and data cables CABB0560 BNC-to-BNC, CBTB0278 triple BNC, CABU0213 USB, 110M8172-012 LEMO-BNC tach Yes
Power 108M4044 AC adapter plus five regional plug types Yes
Carry and protection CBVB0552 soft carry bag, NKST0553 neck strap Bag and strap yes; 100M5828 hard case and DTC70262 dust cover separate
Software System 1 3071/01; Ascent Level 1 108M4051; Ascent Level 2 108M4052 No, licensed separately
Optional hardware BATT0575 battery, KTTC0331 triaxial kit, 108M4069 laser tach, CBL50216 tach cable, 113M5529-01 reflective tape No
Metrology certification No part number published No, charged option at order placement

The asterisk in the datasheet matters. It marks the accelerometer rows where more than one part number is listed and states that those items “have limited regional availability due to certification requirements.” In other words, the kit is standardised, but the sensor supplied inside it can change with the destination market.

Which vb6 specifications matter for route-based collection?

The vb6 records four channels simultaneously with a 24-bit ADC, more than 95 dB dynamic range and Fmax up to 40 kHz. It stores data in dual 1 GB non-volatile flash memories, runs about 10 hours on a charge, and transfers data to a PC over USB or Ethernet for use in Ascent, System 1 or XML-based workflows.

Those numbers define a collector rather than an online protection system. The vb6 is designed to be carried along a route, placed on a bearing housing or casing, and triggered for a repeatable measurement. Its high Fmax and demodulation options support early bearing-fault detection, while the tachometer input allows order-based and phase-related measurements.

Parameter Published value Practical note
Input channels 4, simultaneous sampling CH1 is BNC; CH2 to CH4 are LEMO
ADC 24-bit Applies to all four channels
Dynamic range >95 dB typical at 400 lines Datasheet feature list states 95 dB or greater
Frequency range Fmax up to 40 kHz; Fmin 0 to Fmax Discrete Fmax values from 25 Hz upward
Resolution Up to 12,800 lines 6,400 lines maximum in dual-channel mode; 3,200 in four-channel mode
Sampling rate Not disclosed Waveform lengths of 1,024 to 32,768 samples are published
Storage 1 GB in the feature list; dual 1 GB non-volatile flash in the specification table The second flash holds a mirror copy of the database
Battery Lithium-ion 7.4 V, 5 Ah; 10 hours operating; about 3 hours to charge Backlight times out after 60 seconds
PC connection USB 2.0 host or Ethernet Output formats listed as Ascent and XML
Sensor types Accelerometer, velocity, displacement, current, 4 to 20 mA, voltage output 2.2 mA excitation for IEPE/ICP accelerometers
Tachometer Laser, contact, TTL pulse or Keyphasor 10 RPM to 300,000 RPM; optically isolated input
Environmental IP65, 1.2 m drop test, -10 to 50 degrees C operating 1.2 kg including battery and strap

Two published inconsistencies are worth recording. The feature list says “1 GB memory,” while the specification table describes dual 1 GB flash memories with a mirror copy. The display is listed as 5.7 inches in the vb6 datasheet, whereas the older vb/SCOUT100 brochure describes a 5.5-inch display for the family. For a bill of materials, quote the vb6 datasheet revision you are using rather than the family brochure.

Replacement cabling is a normal consumable in route work. The kit uses four transducer cables, and a damaged cable stops the route until it is replaced. The 200151-40-03-00 accelerometer cable is one of the Bently Nevada cable families stocked for this type of sensor chain, while the 330450-60 MIL-C-5015 connector covers the three-pin interface used on many industrial accelerometers.

Is System 1 included, and which software licence do you need?

No. The vb6 basic kit contains the instrument, sensors and cables, but not a software licence. Bently Nevada lists System 1 as part number 3071/01 and Ascent Level 1 and Level 2 as 108M4051 and 108M4052, all purchased separately. The datasheet names System 1 as compatible software and lists Ascent and XML as output formats.

The datasheet wording is direct: the instrument is offered “in a basic kit with the option to purchase the System 1 software and license separately.” The same page also lists the Ascent licences, so a buyer has to choose the software environment before the instrument is quoted, not after it arrives.

  • System 1, part 3071/01. The datasheet identifies System 1 as the condition-monitoring software the vb6 works with.
  • Ascent Level 1, part 108M4051, and Level 2, part 108M4052. The datasheet lists both as separate software items and names Ascent as an output destination.
  • Communication methods. The vb and SCOUT100 series brochure describes System 1 Evolution links over a USB cable, an Ethernet cable, file mode and Remote Comms over a WAN.
  • What the datasheet does not describe. The exact System 1 import path for vb6 recordings is not set out in the datasheet; it lists Ascent and XML as output formats instead.

Licence mode is another gap. The datasheet gives part numbers but does not state whether each licence is node-locked, seat-based or server-based, nor how many instruments one licence can serve. Those terms belong in the software quotation. The vb6 firmware itself is upgradeable through the PROFLASH system, and the datasheet includes free firmware updates for five years and a five-year hardware warranty.

What is the difference between Standard Test Data and Metrology Certification?

Standard test data is free on request but the datasheet explicitly says it is not evidence of calibration. Metrology certification is a charged service that can be requested when the order is placed. The datasheet does not state an ISO/IEC 17025 accreditation, the issuing laboratory, a certificate validity period or a separate order code for the certificate.

The exact wording appears at the foot of the ordering page and is short enough to quote in full:

“Metrology certification can be requested when an order is placed, however this service is charged. Standard test data can be requested for free, but is not evidence of calibration.”

— Bently Nevada, vb6 Portable Vibration Data Collector Datasheet, 111M7155 Rev. K, page 7.

The same note adds: “To request a periodic metrology calibration, contact Bently Nevada Tech Support.” That sentence covers recalibration after delivery, not the initial certificate. It also stops short of naming the laboratory that performs the work.

Question Standard Test Data Metrology Certification
Cost Free Charged
When to request Can be requested; the datasheet does not say it ships automatically When the order is placed
Calibration evidence Explicitly “not evidence of calibration” Certification of metrology; scope not described
Standard referenced Not disclosed Not disclosed; ISO/IEC 17025 is not named
Issuing laboratory Not disclosed Not disclosed
Validity period Not disclosed Not disclosed
Order code Not disclosed Not disclosed
Contents Not disclosed Not disclosed

For an audit-driven purchase, the safe reading is this: the free document supports acceptance and traceability checks inside the buyer’s own quality system only if that system allows it, while the paid certificate is the item to put on the purchase order when a calibration certificate is mandatory. Do not treat the free data sheet as a substitute, because the manufacturer says it is not one.

ISO/IEC 17025 is the question most buyers ask next, and the datasheet does not answer it. It also does not state whether the certificate is issued by an in-house Baker Hughes laboratory or an external accredited body. Those details have to be confirmed in the quotation, together with the certificate’s validity and the recalibration interval.

What is not disclosed in the public datasheet?

Seven items are absent from the public vb6 datasheet: the sampling rate, the test-data contents, an ISO/IEC 17025 reference, the calibration laboratory, certificate validity, a metrology-certificate order code, and any explicit export-control statement. Each should be confirmed in the quotation rather than inferred from the instrument’s four-channel specification.

This is not a criticism of the document; portable-collector datasheets are written for selection, not for quality-system procurement. The gap becomes a problem when a buyer assumes that a free test record, a paid certificate and an accredited calibration are the same thing. They are not, and the datasheet only defines the first two.

  • Sampling rate. The ADC resolution, Fmax and waveform lengths are published, but the sampling frequency is not. Do not calculate it from Fmax and present the result as a specification.
  • Certificate scope. The datasheet does not list which parameters are calibrated, the uncertainty budget, or the environmental conditions of the calibration.
  • Validity and traceability. No validity period and no traceability chain are stated.
  • Export control. The regional note is tied to certification requirements, not to export licensing. No export-control statement appears in the document.
  • Kit regional variants. The exact sensor supplied under the ACCL0547, ACCL0561 or 200350 alternatives depends on the destination market and is not resolved in the datasheet.

Where does the vb6 fit in a condition-monitoring programme?

Bently Nevada positions the vb6 for on-route and off-route data collection, where a technician walks a route and records bearing, casing or structural vibration at defined points. Official use cases include rolling-element bearing machines, new condition-monitoring programmes, balancing checks and supplemental dynamic data on assets without online monitoring.

The official guidance is specific about the trade-off. Portable collection is periodic, often monthly, and requires a person to move from asset to asset. That makes it suitable for lower-consequence machines and for building a baseline before investing in online monitoring. It is not a substitute for protection systems on critical machinery, where continuous monitoring and trip logic are required.

  • Rolling-element bearing machines. Slow wear modes can be trended from repeat route measurements, with demodulation helping to expose bearing faults early.
  • New programmes. A portable route is a lower-cost entry point than a plant-wide online system and can later be supplemented by wireless or online monitoring.
  • Balancing and alignment support. The vb6 is a collector, not a balancer; Bently Nevada sells separate balancing-capable models, so confirm the function before assuming it is included.
  • Supplemental dynamic data. Routes can add waveforms and spectra to assets that otherwise report only 4 to 20 mA trends.

Vendor language should be read separately from those use cases. The vb and SCOUT100 brochure claims “leading-edge reliability, accuracy and usability” and a “premium return on investment.” Those are marketing positions, not measured performance guarantees. The measurable statements are the datasheet figures: four simultaneous channels, 24-bit conversion, 95 dB dynamic range, 40 kHz Fmax and a five-year hardware warranty.

Fixed sensors still have a role alongside a portable route. The Bently Nevada proximity probe range covers radial shaft measurement on machines that need continuous protection, and vibration transmitters provide a 4 to 20 mA trend for assets that do not justify a full online system. A 330505-01-02-00 low-frequency velocity sensor is an example of the velocity-measurement chain that a portable route may complement.

For sites mixing portables with rack-based monitoring, the Bently Nevada spare parts brand index groups the module, probe and cable families in one place. Our Bently Nevada spare parts guide explains how the rack and sensor families fit together. The Orbit 60 CMM analysis covers the data-path side for plants that also run System 1.

What should be verified before ordering a vb6 or its spares?

Before ordering, confirm four things: whether the route needs four simultaneous channels or only a dual-channel mode, which agency approval matches the plant classification, whether the kit’s alternate sensor part numbers are available in your region, and whether a metrology certificate is required on the purchase order. Software licences and hard cases are separate line items.

These checks prevent the two most common procurement mistakes: ordering an instrument whose approval does not match the area classification, and assuming that the kit price includes System 1 or a calibration certificate. Neither assumption is supported by the datasheet.

  1. Channel count and measurement mode. Specify four-channel capability if the route uses triaxial sensors or multiple points per setup. Remember that four-channel spectra stop at 3,200 lines.
  2. Approval code. The vb6 datasheet lists only approval 01, CSA Class I Division 2. If the plant requires Zone 1 or Division 1, the vb6 is not the correct model and a different Bently Nevada portable should be selected.
  3. Kit regional variant. Ask the distributor which accelerometer part number will ship: ACCL0547, ACCL0561 or 200350. The choice is constrained by certification requirements in the destination market.
  4. Software licence. Add System 1 3071/01 or the required Ascent level as a separate line, and confirm the licence mode and seat count in writing.
  5. Metrology certificate. If a calibration certificate is mandatory, request it when the order is placed and obtain the accreditation, issuing laboratory and validity details in the quotation.
  6. Consumables and case. Add spare transducer cables, a replacement battery and the hard case if the instrument will travel outside a controlled store.

Where do these facts come from?

The facts in this article come from the Baker Hughes vb6 datasheet 111M7155 Rev. K, the vb and SCOUT100 series brochure, Bently Nevada’s portable-collector guidance and the public product pages cited below. No sampling rate, certificate validity, accreditation or metrology order code has been estimated where the manufacturer does not publish one.

Verified 23 November 2026 against the documents above. Where the manufacturer publishes no figure, the article says “not disclosed” rather than estimating one.

Frequently asked questions

Does the vb6 record two channels or four channels at once?

It records four channels at once. All four inputs sample simultaneously through 24-bit converters, and the instrument supports triaxial sensors. Dual-channel operation is a measurement mode with a 6,400-line limit, not the number of physical inputs. Four-channel spectra are limited to 3,200 lines.

Is 111M7155 a part number I can order?

No. 111M7155 is the document number of the vb6 datasheet, currently Rev. K. The orderable instrument code is VB6-01, where 01 is the CSA Class I Division 2 agency approval. Entering 111M7155 into an order system should return a document, not hardware.

Is System 1 included in the vb6 basic kit?

No. The basic kit contains the instrument, four accelerometers, cables, magnetic bases, power accessories and a carry bag. System 1 is listed separately as part 3071/01, and Ascent Level 1 and Level 2 are 108M4051 and 108M4052. All software licences are purchased separately.

Is the free standard test data a calibration certificate?

No. The datasheet says standard test data can be requested free of charge but is not evidence of calibration. A metrology certification is a different, charged service requested when the order is placed. The two documents should not be treated as interchangeable in a quality system.

Which accreditation standard does the metrology certificate follow?

The vb6 datasheet does not name one. ISO/IEC 17025 is not referenced, and the issuing laboratory is not identified. Buyers who require an accredited certificate should ask the distributor to confirm the standard, the laboratory, the uncertainty budget and the validity period in the quotation.

Why do some vb6 kit sensors have two part numbers?

The kit lists ACCL0547, ACCL0561 or 200350 for the accelerometers. The datasheet marks these rows with a note that items with multiple part numbers have limited regional availability due to certification requirements. The distributor should confirm which variant is approved for the destination market before the order is placed.